Finite element modelling and mechanical evaluation of anteversion in femur and femoral prosthesis - Laboratoire de mécanique des structures et des systèmes couplés
Communication Dans Un Congrès Année : 2000

Finite element modelling and mechanical evaluation of anteversion in femur and femoral prosthesis

Résumé

We studied the mechanical effects of anteversion of hip prosthesis, using a geometrical model of the femoral bone derived from computed tomography sections. Linear elastic mechanical behavior was assumed for both cortical and trabecular bone. The results of finite element modeling were studied qualitatively. Stress distribution was evaluated during one-legged stance loading of models with various degrees of anteversion. The undeniable impact of anteversion raises questions about the interpretation of bone remodeling. Anteversion modifies mechanical stresses, most notably those through the femoral shaft; they are rotated around the vertical axis in the same direction as the anteversion. Changes in Von Mises stresses were larger than changes in longitudinal stresses, suggesting that anteversion may have a more significant impact on shear stresses, which may cause failure at the prosthesis-cement interface, particularly toward the middle of the femoral shaft.
Fichier principal
Vignette du fichier
Arafati2000.pdf (633.48 Ko) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-03179609 , version 1 (03-11-2023)

Identifiants

Citer

N. Arafati, J.-Y. Lazennec, Roger Ohayon, G. Saillant. Finite element modelling and mechanical evaluation of anteversion in femur and femoral prosthesis. ASME 2000 International Mechanical Engineering Congress & Exposition, IMECE 2000, Nov 2000, Orlando, Florida, United States. ⟨10.1115/IMECE2000-2284⟩. ⟨hal-03179609⟩
38 Consultations
27 Téléchargements

Altmetric

Partager

More